Features
High-density 1000 V DC multiplexing/scanning (up to 144 two-wire channels in 1U footprint)
Built-in configuration relays expand the individual building blocks
Ideal for hipot, cable breakdown, source measure unit and power supply switching
Extensive signal shielding employed on-board for excellent signal fidelity
Fail-safe interrupts can detect fault condition and automatically open up relays to a default state
|
|
EX1200-2007A Datasheet
48 channel 1000 V Multiplexer
|
Overview
The EX1200-2007A is designed for scanning multiple high-voltage points to a common bus in either 1- or 2- wire configurations. It consists of two individual (1 x 12) 2-wire multiplexers, or dual (1 x 24) 1-wire multiplexers that can be interconnected under program control (via bussing relays) to configure larger multiplexers as required. This reduces the need for external cabling and helps reduce unterminated stub effects.
When switching high voltages, the need for signal shielding becomes critical. The EX1200-2007A has been designed to include large shield planes that reduce crosstalk and voltage spikes to adjacent channels.
A
fail-safe interrupt line is provided on the front panel that can be used when switching high-voltage source/measure units, or power supplies. This interrupt can be programmed to open up all relays automatically in the event of a fault condition and, thus instantly removing power from the user interface.
Up to 144 2-wire channels can be accommodated in a single EX1200 series mainframe for maximum density, or combined with other EX1200 series modules to create a flexible system switch.

|
|
Specifications
|
|
Channel Count |
(1 x 24) 2-wire, Dual (1 x 12) 2-wire, or Dual (1 x 24) 1-wire |
|
Relay Type |
Reed |
|
Maximum Switching Voltage |
1000 V DC |
|
Maximum Switching Current |
1 A |
|
Maximum Carry Current |
2 A |
|
Maximum Switching Power |
25 W dc |
|
Path Resistance |
< 1 Ω |
|
Insulation Resistance |
> 1x107 Ω |
|
Bandwidth (-3 dB) |
15 MHz |
|
Rated Switch Operations |
|
| Mechanical |
1 x 108 |
| Electrical |
1 x 106 at full load |
| Switching Time |
< 1 ms |
| Front Panel Connector |
160-pin DIN (1/2 populated) |
|
|
Ordering Information
|
|
EX1200-2007A |
48 channel 1000 V Multiplexer |
| 52-0109-000 |
100 crimp pins |
| 46-0010-000 |
Crimp tool |
| 46-0011-000 |
Extraction tool |
| 70-0363-504 |
Strain relief bracket kit (includes connector – recommended accessory |
| 70-0363-503 |
Strain relief bracket kit (no connector) |
| 27-0088-160 |
Mating connector |
|
|
|